Switching module

2
00:00:02,398 --> 00:00:29,332
Next, we are going to see how to work with the switching module. If in the switching module we have an Overcurrent error of the board one or board two, we have to open the board of the switching to find its problem.

3
00:00:29,357 --> 00:00:51,009
First, we disconnect the connection between the two sides of the switch, which is this flat,

4
00:00:52,524 --> 00:01:06,103
and then the input power cables that enter the two sides of the switch from the protection circuit.

5
00:01:18,908 --> 00:01:30,642
Then, we open the screws that are attaching the two sides of the Main to the Spacer. Every one of these modules, except for the micro module, is attached to the Main by these screws.

6
00:01:30,667 --> 00:01:34,344
The micro module is attached to the Main by its own pins.

7
00:01:34,369 --> 00:01:44,963
You can easily detach PS,

8
00:01:45,057 --> 00:01:49,534
Fan Controller,

9
00:01:49,651 --> 00:02:05,862
Command and Micro modules from the Main by applying little pressure to the outside.

10
00:02:06,124 --> 00:02:17,698
To detach LAN-GPS, first, the above screws are opened and the Wi-Fi module is detached.

11
00:02:17,847 --> 00:02:36,975
Then, by opening this spacer and a screw in this part the LAN-GPS module is detached.

12
00:02:38,054 --> 00:02:48,147
There is one screw in the end which is in the back of the PS module and two more screws at the bottom of the module.

13
00:03:01,175 --> 00:03:10,472
After all side screws are opened, the Main can be easily detached.

14
00:03:10,497 --> 00:03:17,064
On the Incomplete Main Switching, other than these protective elements, there is a command which,

15
00:04:46,010 --> 00:04:53,844
just like the complete switching side, generates high-frequency pulses and the output of this module is 4 pulses.

16
00:04:53,869 --> 00:05:16,388
We can detach the Incomplete Main Switching by opening the screws around it. Also, it was already mentioned that in addition to the protective modules, there is a command module as well.

17
00:05:16,413 --> 00:05:28,091
The command used in the Complete Switching Side is no different from the one used in the Incomplete Switching Side.

18
00:05:28,193 --> 00:05:34,521
This was mentioned to note that when we have an Overcurrent error on a Switching side,

19
00:05:34,584 --> 00:05:46,966
before opening the Mains, we should swap the commands of the two sides to make sure that the problem is with the command module.

20
00:05:47,076 --> 00:06:00,761
If the error is eliminated the problem is with the command otherwise there is a problem in other parts.

21
00:06:00,802 --> 00:06:10,325
After opening the switching module, it is a little hard to differentiate between the complete and incomplete side.

22
00:06:10,628 --> 00:06:24,979
In the complete switching module side you can see a double pin header. The electricity is AC that reaches the PS module.

23
00:06:25,004 --> 00:06:32,128
Then the PS is electrified so that the other modules are electrified.

24
00:06:32,159 --> 00:06:42,611
There is no such pin header in the incomplete switching side.

25
00:06:42,636 --> 00:06:50,321
Only there is a 3-pin pin header which issues the connection command of the relay of the incomplete side or switch number one.

26
00:06:50,346 --> 00:06:59,971
In the complete side, we have the same triple pin-header which is used for the complete side switch or switch number two.

27
00:06:59,996 --> 00:07:05,507
Also, we have the phase and null that electrifies the PS.

28
00:07:05,532 --> 00:07:15,105
Therefore, the side where this double pin header is located is the AC electricity and because of this, we know that this is the complete side of the device.

29
00:07:15,130 --> 00:07:30,831
When there is an Overcurrent error of board two or board one, as mentioned before,

30
00:07:30,883 --> 00:07:41,135
first we swap the command modules of the two sides. If the problem is solved, the command module is causing the error,

31
00:07:41,127 --> 00:07:49,143
otherwise, we need to open the Main following the mentioned procedure and then move the faulty side up to this level like this.

32
00:07:49,316 --> 00:07:58,882
Here you can check some items to see what part is causing the Overcurrent error.
