To begin discussing Power Control in this device, it is necessary to first understand the design of its power supplies

2
00:00:25,566 --> 00:00:30,465
Since this device is used not only for relay testing but also for testing instruments such as transformers

3
00:00:30,699 --> 00:00:35,565
its power sources have a high burden capability. Depending on the design,

4
00:00:35,899 --> 00:00:44,375
if the equipment under test does not require a high burden, the device’s power is dissipated in the existing MOSFETs and appears as heat.

5
00:00:44,400 --> 00:00:50,266
Therefore, the amplifier burden must be managed based on the connected load in order to prevent temperature rise.

6
00:00:50,398 --> 00:00:55,764
Note that when we talk about the device’s burden, we are referring to the positive and negative voltage limits of the amplifier.

7
00:00:56,165 --> 00:01:03,464
To better understand this, consider a current source capable of injecting 32 A of current and 22 V peak

8
00:01:03,832 --> 00:01:09,600
If the connected load has an impedance of 100 milliohms, injecting 5 A requires only 0.5 V.

9
00:01:10,798 --> 00:01:20,333
Therefore, the remaining 21.5 V appears across the MOSFETs, and based on the power dissipation formula, we can say: P = 21.5 × 5 = 107.5 watts.

10
00:01:20,588 --> 00:01:27,666
This amount of power is dissipated in the MOSFETs and results in increased temperature.

11
00:01:27,691 --> 00:01:33,399
Therefore, the lower the upper and lower voltage limits are, the less power is dissipated on the MOSFET

12
00:01:33,591 --> 00:01:39,424
On the other hand, if the voltage is too low, the current injection may not be performed correctly

13
00:01:39,557 --> 00:01:47,780
Thus, there is an optimal point where the temperature remains low and the upper/lower voltage levels are just sufficient to deliver the requested current.

14
00:01:47,956 --> 00:01:57,446
This adjustment can be done in two ways: automatic and manual. The automatic mode is available only on devices from model 205 - 8 and above

15
00:01:57,688 --> 00:02:01,246
By enabling the Power Control option, the adjustment is performed automatically.

16
00:02:02,045 --> 00:02:06,454
However, because the system must constantly monitor for current spikes at every moment,

17
00:02:06,488 --> 00:02:09,921
it carries out this process conservatively. After enabling this option and starting the test,

18
00:02:10,021 --> 00:02:14,846
you will see the Power Control percentage displayed.

19
00:02:16,320 --> 00:02:21,513
The second method is the manual mode, which is available on all devices, including models 105 and 205

20
00:02:21,538 --> 00:02:24,846
For this method, after designing the test plan,

21
00:02:25,220 --> 00:02:31,046
the software automatically identifies the highest voltage and current values present in the test plan during the Hardware Check step

22
00:02:31,086 --> 00:02:33,913
and uses them to optimize the Power Control

23
00:02:34,380 --> 00:02:38,851
In the highlighted section, the percentage of the amplifier’s voltage

24
00:02:38,951 --> 00:02:46,218
relative to its nominal value is reduced and adjusted as much as possible to minimize it. The user can also manually modify this value.

25
00:02:46,751 --> 00:02:54,913
Ultimately, an optimal value for Power Control is obtained and applied in the software. After running the test

26
00:02:55,234 --> 00:02:59,980
you will see that the calculated value is used during the test for managing the device burden.
